C.R. Bard, Inc. (BASD) -3006260740 PowerPort ClearVUE isp Implantable Port, ChronoFlex Single-Lumen, 6F; Port & Catheter, Implanted, Subcutaneous, Intravascular
|
Back to Search Results |
|
| Catalog Number |
1606062 |
| Medical Device Problem Code |
Device Markings/Labelling Problem (2911)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/27/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On (b)(6) 2026, a patient underwent a port placement procedure using the powerport clearvue isp.Post procedure, it was reported that an x ray was performed to assess for complications and confirmed that the ct marker was inverted; however, the cv port itself was not inverted.A review of 50 cases implanted during august identified one additional case with a similarly inverted marker.It was further noted that the issue was initially identified in a batch manufactured on august 27, and a retrospective review revealed the same issue in a batch manufactured on august 14.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: as the lot number for the device was provided, a review of the device history records will be performed.The device has not been returned to the manufacturer for evaluation.However, photos were provided for review.The investigation of the reported event is currently underway.Section a through f: the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Search Alerts/Recalls
|
|
|