Analysis was performed by remote service personnel which included review of the reported issue, troubleshooting efforts to reproduce the problem, verification of system configuration settings, and review of service documentation.The reported issue was that alarm text was displayed at surveillance station bjhix63sur1, however, no alarm audio was heard.The issue was reproducible.Investigation confirmed that the external speaker was properly connected and had been replaced by the customer without resolving the issue.Further troubleshooting identified that the sound output in the control panel was incorrectly configured to the hrt receiver instead of the external speaker.No error codes or error messages were reported.Software version c.03.06 was installed at the time of the event.Based on the results of the analysis, the product was confirmed to be operating per specifications and the cause of the reported problem was determined to be a configuration issue in the audio output settings.The issue was resolved by reconfiguring the sound output in the control panel to the external speaker.Following the configuration change, the customer verified that alarm audio was functioning correctly and alarms could be heard at the surveillance station.The system met specifications following service and was returned to use.
|