| Model Number |
e601 |
| Medical Device Problem Codes |
High Test Results (2457); Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/29/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The initial reporter received questionable elecsys anti-tg assay results for two patient samples tested on the cobas 6000 e601 module.Patient sample 1
the initial analysis of the patient sample yielded 282.8 iu/ml.The first repeat analysis yielded 20.63 iu/ml.The second repeat analysis yielded 16.85 iu/ml.Patient sample 2
the initial analysis of the patient sample yielded 463.3 iu/ml.The first repeat analysis yielded 18.9 iu/ml.The second repeat analysis yielded 21.5 iu/ml.The repeat results aligned with the patients' expected values.
|
| |
|
Additional Manufacturer Narrative
|
|
The elecsys anti-tg reagent lot number is 87796701, with an expiration date of 31-aug-2026.The field service engineer (fse) inspected the module and identified the ethylene propylene diene monomer (epdm) packing as the cause of the event.The fse replaced this part and verified the module's performance.The investigation determined that a component failure (epdm packing) had caused the event.The investigation determined that the service actions resolved the issue.
|
| |
|