| Catalog Number |
05942861001 |
| Medical Device Problem Code |
Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/24/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
An allegation of questionable blood glucose results was received for a single patient tested consecutively on an accu-chek inform ii meter + rf meter.Testing performed within 15 minutes yielded 323 mg/dl and 145 mg/dl.The lower result was deemed correct.
|
| |
|
Additional Manufacturer Narrative
|
|
The accu-chek inform ii meter + rf meter serial number is (b)(6).The customer stated that the qcs were acceptable within 24 hours of the event.The meter and test strips were requested for investigation, but have not been received.If the product is returned in the future, a follow-up report will be submitted.On a regular basis, accu-chek inform strips of lots currently valid in the market are tested as part of routine retention testing, and the results have passed internal inspection.
|
| |
|
Search Alerts/Recalls
|