| Model Number |
500001002506 |
| Medical Device Problem Code |
Communication or Transmission Problem (2896)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/27/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that when using the bd pyxis¿ medstation¿ es tower device was failed to communicate.The customer stated that this malfunction caused a delay in dispensing medication to patients, but the user was able to get the medications from another station.However, there were no adverse events or injuries reported based on this event.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the complaint history for sn (b)(6)was performed in salesforce which did not locate similar complaint(s) with the same failure mode for this serial number.A review of the device history record for sn (b)(6)was performed from the date of manufacture, 12-jan-2023 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Upon investigation of the actual device used in this incident, it was determined that the device was failed to communicate.A field service engineer found the station offline due to a failed hospital network ethernet port.Established a temporary wireless fidelity (wi-fi) connection to restore communication and coordinated with hospital information technology (it) team regarding permanent network repairs.Verified functionality through (hardware test application) hta and application testing.Confirmed normal operation.The system functioned as intended after the field service engineer repaired the device.
|
| |
|