Model Number SC-1416
Medical Device Problem Codes Unexpected Therapeutic Results (1631); Battery Problem (2885)
Health Effect - Clinical Codes Unspecified Infection (1930); Pain (1994)
Date of Event 06/02/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient with a wavewriter alpha prime was a high energy user and the patient's ipg had reached end of life.The physician found an infection at the pocket site during a revision procedure and decided to explant the ipg including the spinal cord stimulator paddle (scs) lead.
 
Additional Manufacturer Narrative
Block b3: exact date approximated based on the date the manufacturer became aware of the event.Additional suspect medical device component involved in the event: model number/catalog number: sc-8216-50, serial number: (b)(6), batch/lot number: 7079610, model/catalog description: artisan lead 50 cm, unique identifier (udi): (b)(4).Investigation results: all explanted devices were not returned for analysis; therefore, a technical analysis could not be performed.Device history record: it was confirmed that this device met manufacturing specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.Labeling review: review of the instructions for use (ifu) confirmed there was relevant content and sufficient guidance with respect to the circumstances described within this complaint.No updates are required to the ifu as a result of this event.A sterilization record review (srr) was performed, and the review identified no sterilization-related findings that could be reasonably associated with the reported event.Investigation conclusion: all explanted devices were not returned for analysis and the complaint as reported could not be confirmed.Record review revealed no additional information related to all complaints.Therefore, this investigation is unable to determine a probable cause for the complaint and the conclusion code cause not established will be used.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WaveWriter Alpha? Prime 16
Common Device Name
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION CORPORATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
Boston Scientific Corporation
cashel road
clonmel E91 T 862
EI   E91 T862
Manufacturer Contact
ricardo amaro castro
25155 rye canyon loop
valencia, CA 91355
+1(320)800-7570
MDR Report Key50866067
Report Number3006630150-2026-07023
Device Sequence Number13427216
Product Code LGW
UDI-Device Identifier08714729985068
Combination Product (Y/N)N
Initial Reporter StateAR
Initial Reporter CountryUS
PMA/510(K) Number
P030017
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberSC-1416
Device Catalogue NumberSC-1416
Device Lot Number235151
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/26/2026
Initial Report FDA Received Date09/23/2026
Date Device Manufactured01/22/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age82 YR
Patient SexFemale
-
-