| Model Number |
LifeVest WCD 5100 System |
| Medical Device Problem Code |
Circuit Failure (1089)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A us distributor returned a monitor and reported monitor does not power on.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation of monitor has been completed.The reported problem (won't power on) was confirmed due to a defective apps (applications subsystems) board.The root cause for the corrupt programming was unable to be positively identified.There was no adverse event that resulted from the damaged monitor.
|
| |
|