| Model Number |
5076-58 |
| Medical Device Problem Codes |
Capturing Problem (2891); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/08/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that on the pre-discharge device check post implantable pulse generator (ipg) implant, the r waves had decreased and thresholds had increased since implant.A chest x-ray confirmed that the implantable pulse generator (ipg) had moved downwards and the right ventricular (rv) lead dislodged as a result.The ipg and rv lead were repositioned and remain in use. it was considered that the cause was likely weak tissue in the pocket area. no patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|