| Catalog Number |
212760 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/14/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Failed to pass registration, successful after changing reamer shafts.Case type / application: tha 4.1.Update - the device is missing two screws, causing it not to function.
|
| |
|
Additional Manufacturer Narrative
|
|
An event regarding missing component involving a mako reamer handle was reported.The event was confirmed.Method & results:
product evaluation and results: functional inspection confirmed the event.The device is missing two screws, causing it not to function.Clinician review: no medical records were received for review with a clinical consultant.Product history review: a product history review is not required, as no manufacturing specific issue has been alleged.Complaint history review: a complaint history review is not required as there is no patient involvement.Conclusions: no patient was involved and no actual or potential patient harm existed for the alleged event.The alleged failure mode was confirmed.No additional investigation or specific actions are required.If additional information is received then the complaint will be reopened.
|
| |
|