• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medline Industries, LP MEDLINE; PAD, GROUNDING, UNIVERSAL, SPLIT, 9'

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medline Industries, LP MEDLINE; PAD, GROUNDING, UNIVERSAL, SPLIT, 9' Back to Search Results
Catalog Number PAD9165
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Skin Inflammation/ Irritation (4545)
Date of Event 02/24/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "skin irritant x4, attachment breaking from sticking= multiple times.".
 
Additional Manufacturer Narrative
The customer reported "skin irritant x4, attachment breaking from sticking= multiple times," with irritation "where the adhesive was," noticed "immediately" at time of removal, and the skin looked "red and irritated"; treatment was "none" and no serious injury, medical intervention, or follow-up care was reported as "none that i know of, resolved over time." medline quality received 1 ea of pad9165 for evaluation, and physical inspection was conducted with the provided sample.The customer did not provide any evidence of skin irritation caused by the grounding pad and at this time this issue cannot be confirmed; at the same time the issue with the plug has been confirmed, as the plug was damaged.The suspected root cause is a manufacturing issue, and the division will continue to monitor the issue for trending.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
PAD, GROUNDING, UNIVERSAL, SPLIT, 9'
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
pranay abbu
three lakes drive
northfield, IL 60093--275
MDR Report Key50866510
Report Number1417592-2026-48476
Device Sequence Number10715068
Product Code GEI
UDI-Device Identifier20193489082538
UDI-Public20193489082538
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K212093
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberPAD9165
Device Lot Number(10):2712209000
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer03/16/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/24/2023
Initial Report FDA Received Date09/23/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-