(b)(4).Yes
no
malfunction
no
525 (tube)
2199 (no health consequences or impact)
4582 (no clinical signs, symptoms or conditions)
1120 (contamination)
4112 (analysis of information provided by user/third party), 4114 (device not returned)
4256 (malfunction observed without conclusive finding)
4315 (cause not established)
unknown
the account reported finding item is leaking; brown/green stains in tubing, and it was reported that the tubing was discarded and not used for patient care.A photo was received for our investigation; in the provided photo, one can clearly see some staining on the tubing, but the customer has not provided a sample for evaluation.Based on the complaint details, photo evaluation, and our investigation, the issue of "brown/green stains in tubing" can be confirmed and the issue of "item is leaking" cannot be confirmed.At this time, a root cause for the issue of "brown/green staining on tubing" cannot be identified; this complaint has been provided to the vendor for further evaluation, and the supplier quality team will monitor this issue for trending purposes.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|