C.R. Bard, Inc. (BASD) -3006260740 PowerPort Slim Implantable Port, ChronoFlex Single-Lumen, Kit, 8F; Port & Catheter, Implanted, Subcutaneous, Intravascular
|
Back to Search Results |
|
| Catalog Number |
1718000 |
| Medical Device Problem Codes |
Suction Problem (2170); Migration (4003)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/10/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported through the litigation process that, sometime after the port placement, the port allegedly malfunctioned, with difficulty aspirating from the indwelling port, and the port was reported to have migrated.However, no patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H11 - b3.The date received by manufacturer has been used for this field.Manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this lot.Investigation summary: the physical device was not returned for evaluation.No medical records were provided for review.The investigation is inconclusive for the reported port malfunction, suction issue and migration as no objective evidence was provided for review.Based upon the available information, the definitive root cause is unknown.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.Section a through f: the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|