• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Arthrex, Inc. Dissector, 3.0mm x 7cm; SAW BLADE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Arthrex, Inc. Dissector, 3.0mm x 7cm; SAW BLADE Back to Search Results
Model Number Dissector, 3.0mm x 7cm
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 09/02/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an arthroscopic wrist surgery the device produced metal shaving while removing remnants of the sl ligament that prolapsed into the joint.The surgeon tried to remove the metal shaving, but he did not succeed.Some fragments remained in the patient's knee.The surgery was finished successfully with the same device.It was not necessary to switch the surgical technique or do a second surgery.
 
Additional Manufacturer Narrative
Complaint allegation is confirmed by the attached picture, which shows particles of debris on the pictures.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.Per dfu-0215-eo- rev 1 - f - precautions 6.Do not allow the rotating portion of any device to touch any metallic object, such as a cannula or arthroscope, during use.Damage to both devices is likely.Damage to the device can range from a slight distortion or dulling of the blade/burr edge to actual fracture of the tip in vivo.If such contact does occur, inspect the tip.If there are cracks, fractures or dulling, or if there is any other reason to suspect a blade is damaged, replace it immediately.7.Excessive "side-loading" on the device during use does not improve cutting performance and in extreme cases will cause irreversible damage to the device.The most likely cause is attributed to misuse of the device, including exposure to excessive mechanical forces, misaligned insertion, side loading during use and/or allowing the rotating portion to touch any metallic object during use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Dissector, 3.0mm x 7cm
Common Device Name
SAW BLADE
Manufacturer (Section D)
Arthrex, Inc.
1370 creekside boulevard
naples FL 34108 -194
Manufacturer (Section G)
Arthrex, Inc.
1370 creekside boulevard
naples FL 34108 -194
Manufacturer Contact
jared engle
+1(800)933-7001
MDR Report Key50866746
Report Number1220246-2026-04858
Device Sequence Number9223460
Product Code GFA
UDI-Device Identifier00888867211834
Combination Product (Y/N)N
Initial Reporter CountryDA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 09/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberDissector, 3.0mm x 7cm
Device Catalogue NumberAR-9300DS
Device Lot Number15622062
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/22/2026
Initial Report FDA Received Date09/23/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/13/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-