Catalog Number MDS808650
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/05/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that ¿i now have six additional reclining wheelchairs experiencing the same frame issue we¿ve discussed previously.¿.
 
Additional Manufacturer Narrative
The customer stated that reclining wheelchairs were experiencing a frame issue and reported, they were received this way.Medline received one wheelchair and photo; the photo was a close up of the front left corner of a chair, but the model and serial number were not identified.The wheelchair received was used, expanded easily, the seat sits into the seat cradles or saddles without force, and the left rear wheel was bent in slightly.The customer complaint was confirmed for a bent rear wheel; however, the photo attached is not of this chair and bent seat tubing as stated was not confirmed, and the cause was recorded as cannot determine cause with received sample.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This mdr is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
WC,RECL,22",DLA,ELR, 300 LB
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
lauren glaeser
three lakes drive
northfield, IL 60093--275
MDR Report Key50866778
Report Number1417592-2026-48494
Device Sequence Number16131116
Product Code IOR
UDI-Device Identifier00080196303208
UDI-Public00080196303208
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberMDS808650
Device Lot Number73624070001
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer01/08/2026
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/05/2026
Initial Report FDA Received Date09/23/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-