Updated fields: b1, b4, d10, g3, g6, h2, h6(type of investigation, investigation findings, investigation conclusions), h11\
\
corrected field: g2\
\
due to character limitation in block e1: event site name: (b)(6) hospital\
\
kindly revert a1(patient id) and h4 as it should be kept blank.\
\
a getinge field service engineer (fse) evaluated the device and determined that the reel, ac power cord was defective.The power cord was found to have internal damage, resulting in an intermittent open circuit that could not be visually identified.The fse replaced the defective power cord reel assembly.Following the repair, all functional and safety checks were performed in accordance with factory specifications and the device passed all testing.The unit was returned to service.\
no fault logs were available for review.The defective component was discarded by the customer and was not returned for evaluation.
|
|
It was reported by the customer that during routine check, the cardiosave intra-aortic balloon pump (iabp) was not switching on mains supply.There was no patient involved, no harm reported.A getinge field service engineer (fse) was dispatched to evaluate the unit.The fse reported that power cord gone defective and needs to be replaced.After replacing the ac power cord, the problem was solved.All functional and safety checks have been performed to factory specifications, and the unit has been returned to use.
|