| Model Number |
e601 |
| Medical Device Problem Codes |
Low Test Results (2458); Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/27/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
There was an allegation of questionable elecsys troponin t gen 5 stat and elecsys probnp ii stat results for 9 patient samples on a cobas 6000 e601 module.Examples for 2 patients were provided.Patient 1's initial troponin result was 188 ng/l.The repeat result on another e601 module was 483 ng/l.Patient 2's initial probnp result was 116 pg/ml.The repeat result on another e601 module was 543 ng/l.The repeat results were deemed to be correct.The samples were repeated due to an abnormal qc.
|
| |
|
Additional Manufacturer Narrative
|
|
The elecsys troponin t gen 5 stat reagent lot number was 92446500, and the expiration date is 30-apr-2027.The elecsys probnp ii stat reagent lot number is 88096700, and the expiration date is 31-oct-2026.Qc was within range prior to the event.A field service engineer (fse) determined that the pre-wash sipper mechanism was having intermittent issues and replaced it.The investigation determined that the service actions resolved the issue.
|
| |
|