Roche Diagnostics cobas 6000 e601 module; IMMUNOCHEMISTRY ANALYZER Back to Search Results
Model Number e601
Medical Device Problem Codes Low Test Results (2458); Non Reproducible Results (4029)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/27/2026
Type of Reportable Event Malfunction
Event or Problem Description
There was an allegation of questionable elecsys troponin t gen 5 stat and elecsys probnp ii stat results for 9 patient samples on a cobas 6000 e601 module.Examples for 2 patients were provided.Patient 1's initial troponin result was 188 ng/l.The repeat result on another e601 module was 483 ng/l.Patient 2's initial probnp result was 116 pg/ml.The repeat result on another e601 module was 543 ng/l.The repeat results were deemed to be correct.The samples were repeated due to an abnormal qc.
 
Additional Manufacturer Narrative
The elecsys troponin t gen 5 stat reagent lot number was 92446500, and the expiration date is 30-apr-2027.The elecsys probnp ii stat reagent lot number is 88096700, and the expiration date is 31-oct-2026.Qc was within range prior to the event.A field service engineer (fse) determined that the pre-wash sipper mechanism was having intermittent issues and replaced it.The investigation determined that the service actions resolved the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
cobas 6000 e601 module
Common Device Name
IMMUNOCHEMISTRY ANALYZER
Manufacturer (Section D)
Roche Diagnostics
9115 hague road
indianapolis IN 46250 -045
Manufacturer (Section G)
Hitachi High Tech Corp.
882 ichige hitachinaka
ibaraki 312-8 504
JA   312-8504
Manufacturer Contact
natalie catton
9115 hague road
indianapolis, IN 46250
MDR Report Key50866835
Report Number1823260-2026-03902
Device Sequence Number16131329
Product Code JJE
UDI-Device Identifier04015630924714
UDI-Public04015630924714
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K060373
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Numbere601
Device Catalogue Number04745922001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Received by Manufacturer 09/01/2026
Initial Report FDA Received Date09/23/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-