• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAKO Surgical Corp. Handpiece MICS; Orthopedic Stereotaxic Instrument

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAKO Surgical Corp. Handpiece MICS; Orthopedic Stereotaxic Instrument Back to Search Results
Catalog Number 209063
Medical Device Problem Code Difficult to Insert (1316)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/06/2026
Type of Reportable Event Malfunction
Event or Problem Description
Mics handpiece would not lock.Update - collar locking mechanism - dislodged.
 
Additional Manufacturer Narrative
An event regarding locking mechanism failure involving a mako mics was reported.The event was confirmed.Method & results: -product evaluation and results: motor output coupling - dislodged; loose screws on coupling; collar locking mechanism - dislodged; root cause could not be determined -clinician review: no medical records were received for review with a clinical consultant.-product history review: a product history review is not required, as no manufacturing specific issue has been alleged.-complaint history review: a complaint history review is not required as there is no patient involvement.Conclusions: no patient was involved and no actual or potential patient harm existed for the alleged event.The alleged failure mode was confirmed.No additional investigation or specific actions are required.If additional information is received then the complaint will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Handpiece MICS
Common Device Name
Orthopedic Stereotaxic Instrument
Manufacturer (Section D)
MAKO Surgical Corp.
3365 enterprise ave
weston FL 33331
Manufacturer (Section G)
MAKO Surgical Corp.
3365 enterprise ave
weston FL 33331
Manufacturer Contact
sanjana talathi
325 corporate drive
mahwah, NJ 07430
+1(201)831-5000
MDR Report Key50866842
Report Number3005985723-2026-00498
Device Sequence Number13440748
Product Code OLO
UDI-Device Identifier00848486030193
UDI-Public00848486030193
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K172219
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number209063
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer04/17/2026
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 08/24/2026
Initial Report FDA Received Date09/23/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-