The account reported finding a warped mask where the mask would connect to the elbow of circuit and crnas can¿t get the mask to connect to circuit.Medline quality received six units of dynjaa10763a circuit kits; the mask component size 5 were visually inspected for signs of damage, defects or deformities, and finished good production records and finished good pack build were reviewed.The masks have a warped connection port; the root cause could not be determined based on the complaint information and samples provided and based on the complaint details the root cause is considered to be a vendor product issue relating to the component manufacturing process.Our quality team will monitor this issue for trending purposes, and reportability has not changed.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|