| Model Number |
9535-45 |
| Medical Device Problem Code |
High Readings (2459)
|
| Health Effect - Clinical Codes |
Hypoglycemia (1912); Dizziness (2194); Malaise (2359); Diaphoresis (2452); Alteration in Body Temperature (4568)
|
| Date of Event |
09/10/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that an inaccuracy between the continuous glucose monitor (cgm) and the blood glucose (bg) meter occurred.The sensor was inserted into the abdomen on (b)(6) 2026.Data was evaluated and the allegation was confirmed.The probable cause could not be determined via data.The reported glucose values fall within the d zone of the parkes error grid.The data investigation found a difference in values that falls within the c zone of the parkes error grid.No serious or prolonged patient harm or medical intervention was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|