The customer reported receiving a medline blood culture luer lock access device that has a flimsy/bending blood culture, and reported ¿blood culture flimsy bending and break.¿ medline attempted to get more information; no additional information was provided by the customer, and testing was conducted using bcllad alongside bd vacutainer tubes, with 3 vacutainer tubes for each bcllad to simulate a multiple blood draw using a phlebotomy training device with simulated skin and veins.All bcllads functioned as intended, no vacutainer tubes were loose, stuck, or any other defect, and all bcllad, vacutainer tubes, and needles were free from damage or defect; the investigation results were not confirmed.No root cause was identified, no trends were identified for the specified material number and corresponding defect type, and there was no change to the reportability decision associated with this complaint.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|