• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Abiomed, Inc. - 1220648 Impella; Temporary Non-Roller Type Left Heart Support Blood Pump

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Abiomed, Inc. - 1220648 Impella; Temporary Non-Roller Type Left Heart Support Blood Pump Back to Search Results
Catalog Number 0048-0024-JP
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Hemorrhage/Blood Loss/Bleeding (1888); Perforation of Vessels (2135)
Date of Event 09/10/2026
Type of Reportable Event Serious Injury
Event or Problem Description
Impella cp smartassist was inserted via left femoral artery access in a 33-year-old male patient presenting with idiopathic cardiomyopathy.The patient's comorbidities and society for cardiovascular angiography and interventions (scai) shock stage prior to initiation of support were not documented.The clinical course was identified through published literature; therefore, limited information regarding the duration of support and procedural details was available.The publication reported that the patient received heparin anticoagulation with an activated clotting time (act) of 160 seconds prior to the procedure and maintained act values between 160 and 200 seconds during support.Following completion of treatment and impella cp smartassist explant, bleeding was observed at the femoral access site.Evaluation identified a femoral artery dissection with an associated vessel wall defect.The treating physician performed surgical intervention, including arterial clamping and repair of the vessel defect using a 6-0 nesplen patch.Following completion of the repair, clamps were released and hemostasis was achieved.Reported operative time was 15 minutes with an estimated blood loss of 150 cc.The postoperative course was uneventful.The patient was extubated the following day, and contrast-enhanced computed tomography performed on postoperative day 7 demonstrated no abnormalities at the repair site.Sutures were removed on postoperative day 8, and the patient continued medical management without reported sequelae.No additional clinical, hemodynamic, or device performance information was provided within the publication.The femoral artery dissection and bleeding were observed following device explant and required surgical repair.Based on the available information, the contribution of access technique, anticoagulation management, underlying vascular anatomy, or other patient-specific factors cannot be determined.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by abiomed inc or its employees that the report constitutes an admission that the product, abiomed inc, or its employees, caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Impella
Common Device Name
Temporary Non-Roller Type Left Heart Support Blood Pump
Manufacturer (Section D)
Abiomed, Inc. - 1220648
22 cherry hill drive
danvers MA 01923
Manufacturer Contact
derek sammarco
22 cherry hill drive
danvers, MA 01923
+1(781)493-4967
MDR Report Key50866959
Report Number1220648-2026-13237
Device Sequence Number9425383
Product Code OZD
UDI-Device Identifier00813502011371
UDI-Public(01)00813502011371(11)250614(17)270430(10)2026703779
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P140003
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Literature,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number0048-0024-JP
Device Lot Number2026703779
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/09/2026
Initial Report FDA Received Date09/23/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/14/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age33 YR
Patient SexMale
Patient Weight143 LB
-
-