Catalog Number MDS9407
Medical Device Problem Code Calibration Problem (2890)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/05/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "first time that this bp unit was used the hand got stuck on 240.".
 
Additional Manufacturer Narrative
The reported failure was that the bp unit was used and the hand got stuck on 240.Medline received the medline mobile aneroid in like-new condition; upon visual inspection of the sphygmomanometer, the needle was stuck at the 240 mark, then the gauge was calibrated using the screwdriver provided with these units and tested five times by inflating it up to 300mmhg.During this process, medline experienced no issues, the cuff inflated/deflated and normal, the gauge displayed no issues, and overall the sample worked as intended; the complaint was not confirmed.No root cause was identified, and the division has been informed and will continue to monitor trends.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
ANEROID,MOBILE,ADULT CUFF
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
kehkashan khan
three lakes drive
northfield, IL 60093--275
MDR Report Key50867010
Report Number1417592-2026-48518
Device Sequence Number21138070
Product Code DXQ
UDI-Device Identifier40080196771333
UDI-Public40080196771333
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K211084
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberMDS9407
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer06/14/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/05/2023
Initial Report FDA Received Date09/23/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1