"this vmsr submission is part of a retrospective data remediation review.The included events represent 2 malfunction events that medline became aware of during april 1, 2024 through june 30, 2024.Associated lot numbers 220400020471610 and 230500030542410.The quantity associated with the reported issue was 15 devices.The account reported that mds3003 automatic digital blood pressure monitors with wrist cuffs were producing elevated blood pressure readings, including readings in the 200s and upper 100s, on individuals reported to be healthy and young.No samples, photographs, or videos were received for evaluation despite reports that samples had been returned.Therefore, the reported issue could not be confirmed and a definitive root cause could not be determined.Investigation included review of the reported lot numbers, complaint trending, scar activity, and vendor feedback.No confirmed complaints were identified for lot numbers 220400020471610 and 230500030542410, and no adverse trends were identified for the reported defect type.The vendor reviewed device history records associated with the reported lots and reported no similar issues.The vendor further indicated that the exact cause of the reported condition could not be determined without evaluation of the affected devices.The complaint will continue to be monitored through routine trend analysis.0 devices were returned for evaluation.The devices were not labeled for single use.No devices were identified as reprocessed or reused.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.".
|