| Model Number |
72013 |
| Medical Device Problem Code |
Impedance Problem (2950)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/17/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The nalu sales representative (rep) reported that when the physician placed an eic on the lead, that all contacts showed bad impedances.Troubleshooting was done and the lead continued to show bad impedances.The clincial programmer app was restarted and there were now only a couple of contacts that were bad and programing was attempted.The patient was not feeling the stimulation, and the physician repositioned the lead.The physician removed the lead and put it in the other port, and as he did, the top of the esc snapped in two, and a new one was used.The case continued with new lead and a new eic and the case was completed without issue.
|
| |
|
Additional Manufacturer Narrative
|
|
After conducting a post-acquisition, retrospective review of nalu peripheral nerve stimulation complaint records, boston scientific determined that this event meets mdr reporting criteria.The following mdr information is being submitted as part of capa investigation pcar42.All malfunctioning leads were discarded during the procedure and thus are not available for inspection by the firm.There were multiple people handling the devices during the procedure and reports of multiple attempts at connecting and reconnecting the components.Therefore, the root cause of the reported event could not be determined.
|
| |
|