• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. Braun Medical Inc. Infusomat®; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. Braun Medical Inc. Infusomat®; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number 490100
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/05/2026
Type of Reportable Event Malfunction
Event or Problem Description
As reported by the user facility: 30 minutes after etoposide was started rn went into the patient's room and noticed a wet spot underneath the tubing where the chemo was infusing through.The etoposide was running into the y-site of the fluid line (through equashield) closest to the patient.The leak was at this connection point, despite all connection points with the equashield being tightened all the way.Chemo line disconnected from the fluid line and was reconnected directly to the patient to finish the infusion.The spill did not touch the patient.
 
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal number (b)(4).The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Infusomat®
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
B. Braun Medical Inc.
824 12th ave
bethlehem PA 18018
Manufacturer (Section G)
B. Braun Medical Inc.
824 12th ave
bethlehem PA 18018
Manufacturer Contact
elena reifenstein
824 12th ave
bethlehem, PA 18018
+1(833)425-1464
MDR Report Key50867042
Report Number2521402-2026-00753
Device Sequence Number10715081
Product Code FPA
UDI-Device Identifier04046964186080
UDI-Public(01)04046964186080
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K142036
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number490100
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/17/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/08/2026
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date09/23/2026
Supplement Report FDA Received Date09/25/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-