The customer stated that "the bed arrived with a damaged motor," and indicated that the outer carton was damaged.Medline reviewed the photo evidence provided by the customer; a physical sample was not provided, and only a visual review could be conducted.Investigation results were confirmed: the product appeared to be in new condition as the unit was received this way and the control box housing appeared to be fractured.The cause was updated to dc:cause traced to transport/storage, osi: shipping damage, medline facility; no trends were identified, no product changes were identified that could have contributed to the alleged product failure, and the division will monitor complaints through quarterly trending.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|