The customer reported multiple iv admin sets leaking and that they are having to readminister iv's during surgery.A photo was received for the investigation; the picture highlights one of the y-sites on the device with a blue drape behind it, and liquid is visible on the blue drape, but the picture does not make it clear where from the device the leak occurred.Based on the complaint details, photo evaluation, and investigation, the complaint is confirmed.Despite attempts to gain clarity on how the product was used and where the leak originated from on the device, answers were not received from the customer; therefore, the root cause for the reported issue remains undetermined, and the division will monitor this issue for trending purposes.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|