The customer reported that on (b)(6) 2025 during a paracentesis the tubing started to kink when it was under pressure with suction, temporarily impacted the flow of the fluid being removed from the patient, and the procedure was still able to be completed in its entirety with no injury or impact to the patient.A photo was received for the investigation, finished good production records were reviewed, and the finished good pack build was reviewed.Per evaluation of the photos received, the photo shows the tubing is twisted/starting to kink, confirming the reported issue.This item is received from the vendor and medline does not alter the component before placing it; however, as no physical sample was received for evaluation, medline cannot definitively confirm the root cause, and the complaint has been provided to the vendor for further evaluation.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|