| Model Number |
7740 |
| Medical Device Problem Codes |
Pacing Problem (1439); Under-Sensing (1661)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/09/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that this right atrial (ra) lead exhibited undersensing that led to overpacing.Technical services (ts) was consulted and found that the atrial pacing percentage had increased over the previous month.Ts believed the root cause was an incomplete atrioventricular search interval.No adverse patient effects were reported.This ra lead remains in service.
|
| |
|
Additional Manufacturer Narrative
|
|
This product investigation is still in progress.This report will be updated when product investigation is complete.If pertinent information is provided in the future, a supplemental report will be submitted.
|
| |
|