| Catalog Number |
MDT2168284 |
| Medical Device Problem Code |
Contamination (1120)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/31/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that "upon opening the product had loose lint on the towels.These are supposed to be lintless so they could not be used.".
|
| |
|
Additional Manufacturer Narrative
|
|
The reported issue was that upon opening the product had loose lint on the towels, and these are supposed to be lintless so they could not be used.A sample of the product was received from the customer in an open condition, and 6 months of complaint history for this item was evaluated with no similar complaints reported for this issue of foreign particulates.From the sample received, the issue reported was confirmed; however, it could not be established that the root cause was directly related to the manufacture of the product.At this time, this is not a trending issue, and the divisional quality representative will continue to monitor this issue for tracking and trending purposes during quarterly trending.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|
| |
|
Search Alerts/Recalls
|
|
|