Model Number AFR-00001
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Heart Block (4444)
Date of Event 08/26/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during a cardiac ablation procedure for treatment of atrial fibrillation with the sphere 9 catheter, pulsed field ablation and approximately eight radiofrequency ablation lesions were performed.Pulmonary vein isolation, posterior wall ablation, and an anterior mitral line were created.The patient converted from atrial fibrillation to sinus rhythm during pulmonary vein isolation, and atrial flutter was induced.Complete type 3 heart block occurred during radiofrequency ablation on the anterior wall at the anteroinferior side of the mitral valve.The patient developed an underlying junctional rhythm.The ablation target location and proximity to the conduction system were not confirmed with imaging before ablation.Dopamine was administered, and a temporary pacing wire was placed.Cardiac conduction recovered overnight.The procedure was aborted.No energy-delivery or electrogram issues were observed, and no issues were identified with the sphere 9 catheter.The patient was doing well, and was being discharged.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Sphere-9
Common Device Name
PERCUTANEOUS CARDIAC ABLATION CATHETER FOR TREATMENT OF ATRIAL FIBRILLATION WITH
Manufacturer (Section D)
MEDTRONIC, INC.
880 technology park drive
billerica MA 01821
Manufacturer (Section G)
MEDTRONIC, INC.
880 technology park drive
billerica MA 01821
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
+353(091)708096
MDR Report Key50867370
Report Number3012520654-2026-00520
Device Sequence Number10661386
Product Code QZI
UDI-Device Identifier00763000993269
UDI-Public(01)00763000993269(17)270618(10)0233810333
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
P240013
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberAFR-00001
Device Catalogue NumberAFR-00001
Device Lot Number0233810333
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/26/2026
Initial Report FDA Received Date09/23/2026
Date Device Manufactured05/19/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age74 YR
Patient SexFemale
Patient Weight187 LB
-
-