The hospital reported that vasoview hemopro 3 was done being used but still plugged in.It started smoking and was burning like left alone.They noticed the jaws of the hemopro 3 was burning when it was in neutral position inside the leg.She took it out immediately and circulating rn unplugged the cord and plugged it back in.As soon as the device was plugged in, the jaw was burning on its own.At this point, we decided the device was not safe to use and opened a new vasoview kit.There was nothing between the jaws that could be associated with the smoking of the tool, such as leftover gauze, fatty tissue, char, or blood.The troubleshooting steps taken were to unplug and replug the tool and then open and use a new kit.The smoke was coming from the jaws.The jaw was observed to be burning.No damage was noticed.The activation toggle was in the neutral state when the smoking and burning of the tool was noticed.The toggle was positioned inside the harvester while the tool was being held by the pa.The harvesting tool was placed in its original box and put away for the representative to pick up.There was no patient harm.An audible beep from the power supply was heard when the smoke was noticed.No injuries.
|