The internal complaint file #(b)(4) has been logged for this incident for traceability.The ri-2 and disposable have not yet been returned to belmont for investigation.It was reported to belmont that there was a smoking incident with the ri-2.Additional information was provided on 8/17 that the patient involved expired.The device did not cause or contribute to the death as treatment was continued with a second device.Belmont is filing a report as per procedure.Belmont was informed the unit and disposable set are quarantining at the site in biomedical engineering.Belmont has offered retraining to the customer, however the site has declined.Additional information such as a photo of the disposable has been requested from the site but has yet to be provided.The lot number is unknown.Belmont has requested the device and disposable set to be returned for evaluation once they are released from quarentine.Certain infusates such as calcium, and platelets should not be used, as they can compromise the anticoagulants in citrated blood and promote clotting, which can lead to clot formation inside the heat exchanger and block blood flow.If this occurs, the stainless steel rings inside the heat exchanger may become overheated.In case of overheat alarm, the system stops pumping and heating, closes off the line to the patient, sounds an audible alarm, and displays an alarm message with instructions for corrective measure.Other methods can be used to infuse the patient.Without results of the device and disposable investigation, no conclusions can be drawn at this time.A follow-up report will be submitted once the investigation is complete.
|