A review of the complaint history for sn (b)(6) was performed in salesforce which did not locate similar complaint(s) with the same failure mode for this serial number.A review of the device history record for sn (b)(6) was performed from the date of manufacture, 30-jul-2025 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Upon investigation of the actual device used in this incident, it was determined that the bd external inbound processor service becoming unresponsive, preventing patient adt (admit discharge transfer) messages from processing and causing patient data to stop crossing over to the pyxis system.A technical support specialist (tss) restarted the service, which allowed the backlog of approximately 2,400 messages to process successfully.Patient data synchronization resumed normally, no further issues were observed, and the customer confirmed that the outage was resolved.The system functioned as intended after the technical support specialist troubleshot the device.
|