Details of complaint:
the biomedical engineer (bme) reported the bedside monitor (bsm) is intermittently giving inaccurate non-invasive blood pressure (nibp) readings, and the blood pressure pump sounds very weak.Customer indicated that multiple function tests were performed and the nibp pump is failing when compared to a known good unit.No patient harm was reported.Investigation summary:
the reported issue was duplicated.Evaluation shows that the device failed the nibp check, and the root cause of the reported issue is due to nibp pump malfunction.No further investigation will be performed.Although there was a patient involved, there were no reports of injury, no harm, nor any adverse events, due to the reported issue.Trending will continue to be monitored for this device, incidents, issues, and causes.
|