Model Number BSM-1733
Medical Device Problem Code Incorrect Measurement (1383)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/03/2026
Type of Reportable Event Malfunction
Event or Problem Description
The biomedical engineer (bme) reported the bedside monitor (bsm) is intermittently giving inaccurate non-invasive blood pressure (nibp) readings, and the blood pressure pump sounds very weak.Customer indicated that multiple function tests were performed and the nibp pump is failing when compared to a known good unit.No patient harm was reported.
 
Additional Manufacturer Narrative
Details of complaint: the biomedical engineer (bme) reported the bedside monitor (bsm) is intermittently giving inaccurate non-invasive blood pressure (nibp) readings, and the blood pressure pump sounds very weak.Customer indicated that multiple function tests were performed and the nibp pump is failing when compared to a known good unit.No patient harm was reported.Investigation summary: the reported issue was duplicated.Evaluation shows that the device failed the nibp check, and the root cause of the reported issue is due to nibp pump malfunction.No further investigation will be performed.Although there was a patient involved, there were no reports of injury, no harm, nor any adverse events, due to the reported issue.Trending will continue to be monitored for this device, incidents, issues, and causes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Life Scope PT
Common Device Name
Bedside Monitor
Manufacturer (Section D)
Nihon Kohden Corporation
attn: shama mooman
1-31-4 nishiochia
shinjuku-ku, tokyo 161-8 560
JA  161-8560
Manufacturer (Section G)
Nihon Kohden Corporation
attn: shama mooman
1-1 tajino
tomioka city, gunma 370-2 314
JA   370-2314
Manufacturer Contact
shama mooman
seibu bldg 2, 4th floor 1-11-2
kusunokidai tokorozawa, saitama 359-8-580
JA   359-8580
MDR Report Key50867930
Report Number8030229-2026-07150
Device Sequence Number10673968
Product Code MHX
UDI-Device Identifier04931921111833
UDI-Public(01)04931921111833
Combination Product (Y/N)N
Initial Reporter StateHI
Initial Reporter CountryUS
PMA/510(K) Number
K220976
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 09/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberBSM-1733
Device Catalogue NumberBSM-1733A
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/18/2026
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/03/2026
Initial Report FDA Received Date09/23/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/18/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-