• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Nalu Medical, Inc Nalu Neurostimulation System; Peripheral Nerve Stimulator

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Nalu Medical, Inc Nalu Neurostimulation System; Peripheral Nerve Stimulator Back to Search Results
Model Number 71005
Medical Device Problem Code Impedance Problem (2950)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/25/2023
Type of Reportable Event Malfunction
Event or Problem Description
The nalu sales representative (rep) reported impedance issues during implant procedure as well as during initial activation and programming, so she tried to utilize tight bipoles as close to the psp algorithm suggested locations on lead b as possible.The patient seemed to have decent coverage and went out of town after that programming.No equipment was replaced and all components remained implanted.There was no reported patient harm or injury.
 
Additional Manufacturer Narrative
After conducting a post-acquisition, retrospective review of nalu peripheral nerve stimulation complaint records, boston scientific determined that this event meets mdr reporting criteria.The following mdr information is being submitted as part of capa investigation pcar42.The reported device remains implanted, and the rep has not heard any complaints from the patient at this time.Therefore, the root cause of the reported event could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Nalu Neurostimulation System
Common Device Name
Peripheral Nerve Stimulator
Manufacturer (Section D)
Nalu Medical, Inc
2320 faraday ave
suite 100
carlsbad CA 92008 -724
Manufacturer (Section G)
Nalu Medical, Inc
2320 faraday ave
suite 100
carlsbad CA 92008 -724
Manufacturer Contact
rebecca hess
2320 faraday ave
suite 100
carlsbad, CA 92008--724
MDR Report Key50867972
Report Number3015425075-2026-00660
Device Sequence Number15797766
Product Code GZF
UDI-Device Identifier00812537036137
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K183579
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date02/09/2026
Device Model Number71005
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/25/2023
Initial Report FDA Received Date09/23/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/09/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-