Catalog Number K1186N22S
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/05/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "the arm did not stop in place" and "the hardware that holds the left armrest and stops it from going all the way back has fractured.".
 
Additional Manufacturer Narrative
The customer reported that the arm did not stop in place and that the hardware that holds the left armrest and stops it from going all the way back has fractured; no injuries reported.A physical sample was not provided, and a photo depicting the defect was provided; after reviewing the photo, it was discovered that the left side, armrest rear receiver or pivot bracket appears to have fractured and the extended piece that stops the armrest from pivoting too far appears to have fractured off.The investigation results were confirmed.Given the evidence provided in the photo, as well as the incident report, the root cause of this issue is unknown; no trends were identified for the specified material number and its corresponding defect type, and quality control inspection records indicated the serial number passed inspection.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
WC,K1,18",SB DLA,SA, 300 LB
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
rizwan shaik
three lakes drive
northfield, IL 60093--275
MDR Report Key50867981
Report Number1417592-2026-48658
Device Sequence Number9425410
Product Code IOR
UDI-Device Identifier00193489001061
UDI-Public00193489001061
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K990463
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberK1186N22S
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/05/2026
Initial Report FDA Received Date09/23/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/01/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1