Catalog Number 190766
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code Hemorrhage/Blood Loss/Bleeding (1888)
Date of Event 09/02/2026
Type of Reportable Event Malfunction
Event or Problem Description
A user facility¿s registered nurse (rn) reported that a fme 2008t hemodialysis (hd) machine leaked from the blood pump segment two hours and fifty minutes after the initiation of a patient¿s hemodialysis (hd) treatment.The machine, a fme 2008t machine, alarmed appropriately with an air detector alarm.A fme dialyzer was in use.The blood pump segment of the bloodline had a slit in it.The patient¿s estimated blood loss (ebl) was approximately 5ml.The patient was not restarted because they only had ten minutes left in their treatment.The machine was removed from service following the incident.The biomedical technician (bmt) was not able to find any issues with the blood pump; however, he did replace the blood pump and clip as a precaution because this is the second incident of this nature on this machine.There was no patient injury or medical intervention required due to the reported event.The complaint device is not available to be returned to the manufacturer for evaluation.
 
Additional Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2008T HEMODIALYSIS SYSTEM W/BIBAG
Common Device Name
Dialyzer, High Permeability With Or Without Sealed Dialysate System
Manufacturer (Section D)
ERIKA DE REYNOSA, S.A. DE C.V.
mike allen #1331
parque industrial reynosa
reynosa 88780
MX  88780
Manufacturer (Section G)
ERIKA DE REYNOSA, S.A. DE C.V.
director, quality systems
900 w zaragosa drive suite d
pharr TX 78577
Manufacturer Contact
jessica trujillo
920 winter st
waltham, MA 02451
+1(617)417-5172
MDR Report Key50868087
Report Number0008030665-2026-02543
Device Sequence Number16131124
Product Code KDI
UDI-Device Identifier00840861100910
UDI-Public00840861100910
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K231125
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 09/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number190766
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Device AgeMO
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/02/2026
Initial Report FDA Received Date09/23/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/29/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
FME Bloodlines; FME Dialyzer
Patient SexFemale
Patient Weight162 LB
-
-