Bard Peripheral Vascular, Inc. Maxcore; Biopsy Instrument Back to Search Results
Catalog Number MC1825
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/17/2026
Type of Reportable Event Malfunction
Event or Problem Description
On (b)(6) 2026, a patient underwent a kidney biopsy procedure using a maxcore instrument.During the procedure, the cannula failed to retract automatically, making it impossible to obtain a tissue sample.There was no reported patient injury.
 
Additional Manufacturer Narrative
H11: as the lot number for the device was not provided, a review of the device history records could not be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) are adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow up report will be submitted as applicable.D4 medical device lot # pekz0661 was reported, however, this is not a lot # manufactured for the reported catalog # mc1825.Section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Maxcore
Common Device Name
Biopsy Instrument
Manufacturer (Section D)
Bard Peripheral Vascular, Inc.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
Bard Reynosa S.A. de C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
michael gaffney
800 w. rio salado pkwy
tempe, AZ 85281
+1(480)303-2689
MDR Report Key50868110
Report Number2020394-2026-02800
Device Sequence Number16131243
Product Code KNW
UDI-Device Identifier00801741084492
UDI-Public(01)00801741084492
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K133948
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/07/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberMC1825
Device Lot NumberUnknown
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/01/2026
Initial Report FDA Received Date09/23/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-