| Catalog Number |
MZ1000 CHINA |
| Medical Device Problem Codes |
Leak/Splash (1354); Connection Problem (2900)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/05/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported, maxzero, connection issues, causing leak.The following was provided by the initial reporter: during the procedure to insert an iv cannula on (b)(6) 2026, the nurse connected a clear, cannula-free connector to the patient.The connector did not lock securely, causing fluid leakage.The nurse immediately replaced it with a new clear, cannula-free connector.
|
| |
|
Event or Problem Description
|
|
No additional information.
|
| |
|
Additional Manufacturer Narrative
|
|
H.3.A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
|
| |
|
Additional Manufacturer Narrative
|
|
Annex f code changed from f26 to f2301.Investigation result:
the complaint sample was not available for investigation; however, the customer confirmed that the sample mz1000 china was from lot 25095139.The customer reported that the connector did not lock securely, causing fluid leakage.No further information was available to assist the investigation.The details of the feedback were forwarded to the manufacturing site for investigation.A review of the production records for lot 25095139 did not identify any in-process testing failures or quality deviations that may have caused or contributed to a report of this nature.The root cause of the reported experience could not be determined because no sample was received for examination.Without a sample, it was not possible to determine whether a manufacturing defect could have caused or contributed to the reported event.A review of available complaint history for the relevant batch and associated defect code was conducted as part of the investigation.The information reviewed did not change the conclusions reached during the investigation.Complaint data continues to be monitored through established quality system processes, as appropriate.What does this mean for me?
based on the information available from the investigation, the root cause of the reported event could not be determined because the complaint sample was not available for examination.Without a sample, it was not possible to conclusively link the feedback to a specific failure mode or determine whether a manufacturing defect could have caused or contributed to the event.What happens next?
if you require product replacement or credit consideration, please contact your bd representative or customer care team for assistance.
|
| |
|