The customer reported that after the products were "returned from the field and went through the cleaning procedure" they developed "poor and intermittent suctioning" and that the incidents did not occur while using on a patient or individual.A sample was received for investigation; the sample was visually inspected with no external physical damage observed, plugged into a power source and turned on, and the control dial was turned to the maximum setting, where the pressure was observed to be 600 mm hg and met product specifications.Medline could not confirm the issue of the device having poor suction and could not confirm the product failed to meet specifications.The cause was not confirmed with the received sample, and the quality team will monitor this issue for trending purposes.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|