Olympus Winter & Ibe GmbH Video telescope "ENDOEYE HD II", 5.4 mm, 30°, autoclavable; rigid video laparoscope
|
Back to Search Results |
|
| Model Number |
WA50052A |
| Medical Device Problem Codes |
Break (1069); Contamination (1120); Poor Quality Image (1408); Dent in Material (2526); Output Problem (3005); Scratched Material (3020)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported the rigid video laparoscope had a green image during reprocessing.During the initial evaluation of the device, fiber breakage, minor scratches on the frame, dents and scratches on the outer tube, failed light transmission, and minor stains under the light guide cover glass were observed.No patient involvement was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Date of event is unknown.Additional information will be provided if available.The final investigation has been completed.The failure reported by the customer and identified during the inspection have been confirmed.A root cause could not be identified.Based on the results of the investigation, the most probable cause of the reported malfunction could not be determined.Additionally, the most probable cause of the malfunctions, fiber breakage, minor scratches on the frame, dents and scratches on the outer tube, and failed light transmission, was traced to component failure, and the cause of the minor stains under the light guide cover glass could not be established.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|