It was reported to boston scientific corporation that an axios stent and electrocautery enhanced delivery system was intended to be implanted in the gallbladder to treat cholecystitis during an endoscopic ultrasound gallbladder drainage procedure performed on (b)(6)2026.During the procedure, the physician reported that the stent failed to deploy.The procedure was able to be completed by using another of the same device.There were no patient complications reported as a result of this event.Note: it was reported that the entire handle was rotated.However, the hot axios stent and delivery system instructions for use state "rotate the winged luer lock clockwise to secure the delivery system handle to the echoendoscope." the physician did not follow the steps cited in the instructions for use (ifu).Note: it was reported that the size of the scope used was 3.2 mm working channel.However, per the axios stent and electrocautery- enhanced delivery system instructions for use, the working channel of the scope should be 3.7mm or larger.The stent is not compatible with a scope with a working channel size of 3.2 mm.
|