Model Number 72404127
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Erosion (1750); Urinary Incontinence (4572)
Date of Event 08/01/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient with this artificial urinary sphincter (aus) experienced recurrent urinary incontinence and urethral erosion, and the cuff was previously explanted.The erosion was diagnosed by cystoscopy.A surgical procedure was performed in which the pump was explanted, while the pressure regulating balloon was drained and retained.There were no further patient complications reported.
 
Additional Manufacturer Narrative
The device is not available for analysis; therefore, no physical or visual analysis of the product could be performed.There was no report of a device performance allegation.A review of the device instructions for use (ifu) was completed and did not reveal any evidence of device misuse, off label use, or failure to follow instructions.The reported patient symptom(s) are known risk associated with this device type and indicated as such in the instructions for use.Based on the information available, the conclusion code of known inherent risk of device was assigned to this investigation.This record was identified during a retrospective review of complaint events; an electronic medical device report (emdr) was not previously submitted per applicable reporting criteria.Boston scientific has opened a nonconforming events and prevention investigation (b)(4) to determine the potential cause for why an emdr was not previously submitted to the food and drug administration (fda) in accordance with boston scientific's established medical device reporting process and fda adverse event reporting requirements.Information gained through these investigation efforts will be utilized to take action(s), as necessary.Model number/catalog number: 72400024, serial number: n/a, batch/lot number: 685859007, model/catalog description: balloon fgs 61-70 cm, unique identifier (udi)#: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS 800 Artificial Urinary Sphincter
Common Device Name
Device incontinence mechanical/hydraulic
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
Boston Scientific Corporation
300 boston scientific way
marlborough MA 01752
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key50868660
Report Number2124215-2026-50643
Device Sequence Number18814103
Product Code EZY
UDI-Device Identifier00878953003078
Combination Product (Y/N)Y
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
P000053
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date11/13/2012
Device Model Number72404127
Device Catalogue Number72404127
Device Lot Number682272015
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/26/2026
Initial Report FDA Received Date09/23/2026
Date Device Manufactured12/02/2010
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age84 YR
Patient SexMale
-
-