It was reported that the patient with this artificial urinary sphincter (aus) experienced recurrent urinary incontinence and urethral erosion, and the cuff was previously explanted.The erosion was diagnosed by cystoscopy.A surgical procedure was performed in which the pump was explanted, while the pressure regulating balloon was drained and retained.There were no further patient complications reported.
|
|
The device is not available for analysis; therefore, no physical or visual analysis of the product could be performed.There was no report of a device performance allegation.A review of the device instructions for use (ifu) was completed and did not reveal any evidence of device misuse, off label use, or failure to follow instructions.The reported patient symptom(s) are known risk associated with this device type and indicated as such in the instructions for use.Based on the information available, the conclusion code of known inherent risk of device was assigned to this investigation.This record was identified during a retrospective review of complaint events; an electronic medical device report (emdr) was not previously submitted per applicable reporting criteria.Boston scientific has opened a nonconforming events and prevention investigation (b)(4) to determine the potential cause for why an emdr was not previously submitted to the food and drug administration (fda) in accordance with boston scientific's established medical device reporting process and fda adverse event reporting requirements.Information gained through these investigation efforts will be utilized to take action(s), as necessary.Model number/catalog number: 72400024,
serial number: n/a,
batch/lot number: 685859007,
model/catalog description: balloon fgs 61-70 cm,
unique identifier (udi)#: (b)(4).
|