• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medline Industries, LP MEDLINE; METER,GLUCOSE,EVENCARE G3,PROF.USE ONLY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medline Industries, LP MEDLINE; METER,GLUCOSE,EVENCARE G3,PROF.USE ONLY Back to Search Results
Catalog Number MPH3540
Medical Device Problem Code Low Readings (2460)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/05/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that 1ea of their (b)(6) glucometers is reading low control solution out of range of what is printed on the strip bottle.I asked customer to try again while on the phone and it read 33 while the test strip bottle states 39-69.
 
Additional Manufacturer Narrative
The customer reported a glucometer was reading low control solution out of range of what is printed on the strip bottle, and no medical intervention, serious injury, or follow-up care has been reported.A sample(s) was not received for investigation; investigation included review of recent complaint trends/scar activity for the last three trend reports, and per vendor, through internal investigation including dhr review and inspection records review, no abnormalities were found.The investigation results were unconfirmed, and the complaint was not confirmed due to the fact that this issue can only be confirmed by further investigating the sample or a provided photo; the recorded cause was sample required to determine cause.No trends were identified for the specified material number and its corresponding defect type, and division will continue to monitor trends.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
METER,GLUCOSE,EVENCARE G3,PROF.USE ONLY
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
kumar vigrahala
three lakes drive
northfield, IL 60093--275
MDR Report Key50868675
Report Number1417592-2026-48734
Device Sequence Number16131345
Product Code CGA
UDI-Device Identifier10884389831199
UDI-Public10884389831199
Combination Product (Y/N)N
Initial Reporter StateMS
Initial Reporter CountryUS
PMA/510(K) Number
K120448
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberMPH3540
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/05/2025
Initial Report FDA Received Date09/23/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-