| Model Number |
TJF-Q180V |
| Medical Device Problem Codes |
Corroded (1131); Material Discolored (1170); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/26/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported that the duodenovideoscope experienced distal tip loss during reprocessing.In addition, during the initial evaluation of the returned device, chipping of the adhesive around the light guide (lg) lens and objective lens, as well as discoloration inside the light guide (lg) lens was observed.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The final investigation has been completed.The failure reported by the customer and identified during the inspection have been confirmed.A root cause could not be identified.Based on the results of the investigation, the most probable cause of the reported malfunction was due to pinhole on channel tube.Additionally, malfunction identified during the initial evaluation is traced to component failure.Olympus will continue to monitor the field performance of this device and evaluate any additional information received through post-market surveillance activities.
|
| |
|