• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medline Industries, LP MEDLINE; NEEDLE,HYPODERM,SAFETY, 22GX1.5

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medline Industries, LP MEDLINE; NEEDLE,HYPODERM,SAFETY, 22GX1.5 Back to Search Results
Catalog Number SSN100227
Medical Device Problem Codes Detachment of Device or Device Component (2907); Unintended Movement (3026)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/07/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the item has the same type of flip safety feature that falls easily and prevents provider from giving the injection until it can be moved out of the way; safety cover is loose and flips, nearly covering the needle.
 
Additional Manufacturer Narrative
It was reported that the item have the same type of flip safety feature that falls easily and prevents provider from giving the injection until it can be moved out of the way; safety cover is loose and flips, nearly covering the needle.Medline requested additional information, but there were no samples, photos, videos, serial numbers, lot numbers, or other critical details provided, even after follow-ups with the customer.The complaint could not be confirmed, and medline was unable to confirm the complaint issue or that the product failed to meet specifications.Without a sample received it is difficult to determine a true root cause for the reported issue; if a sample, photo, and/or additional information becomes available, this complaint will be reopened to be further investigated.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
NEEDLE,HYPODERM,SAFETY, 22GX1.5
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
herchelle teodoro
three lakes drive
northfield, IL 60093--275
MDR Report Key50868833
Report Number1417592-2026-48763
Device Sequence Number18469972
Product Code FMF
UDI-Device Identifier30193489031946
UDI-Public30193489031946
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
K213054
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberSSN100227
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/07/2023
Initial Report FDA Received Date09/23/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-