It was reported that a patient underwent a paroxysmal atrial fibrillation ablation with an octaray mapping catheter and the patient experienced a pericardial effusion that required heparin reversed and a pericardiocentesis was performed.Slightly after the transseptal was performed, the patient had a drop in blood pressure and a pericardial effusion was confirmed with intracardiac echocardiography (ice).The medical intervention performed was that heparin was reversed and a pericardiocentesis was performed.The procedure was aborted.The last known patient status was stable.The physician believed that the pericardial occurred during the transseptal process.The octaray mapping catheter had been advanced into the body shortly after transseptal access and that a left atrial pre-map was had been performed using the octaray mapping catheter.No ablation had been performed.The smart touch sf de catheter was not advanced into the body or used for ablation.The carto 3 system was used for mapping.Multiple attempts have been made to obtain clarification of this complaint.However, no further information has been made available.
|
|
The bwi product analysis lab received the device for evaluation 14-sep-2026.The analysis has begun but is not completed at this time.When the investigational analysis has been completed, a supplemental 3500a report will be submitted.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's reference number: (b)(4).
|