Abbott Laboratories ARCHITECT c16000 Processing Module; ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE
|
Back to Search Results |
|
| Catalog Number |
03L77-01 |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/03/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer observed falsely depressed sodium results on the architect c16000 processing module for one patient.The following results were provided:
sid: (b)(6), initial sodium result= 115.4 mmol/l; repeat result on au5800 platform= 136 mmol/l; repeat on architect after recalibration= 137.6 mmol/l.Reference ranges:
abbott system: 136 - 145 mmol/l,
au5800: 135-145 mmol/l.There was no impact to patient management reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Section a patient information: refer to section b5 for complete patient information.The field service representative (fsr) performed troubleshooting on the architect c16000 processing module, serial number: (b)(6) and observed that the washer, cuvette 16 was clogged.The washer, cuvette 16 (rohs) was cleaned and deemed the cause of the issue.The module function was verified with a control/precision run.The instrument service history for the (b)(6) verifies no subsequent issues have been reported related to discrepant patient results after the current complaint was reported.A review of trending data associated with the washer, cuvette 16 (rohs) did not identify any trends.A review of tracking and trending did not identify any trends for the architect c16000 processing module with regard to the customer reported event.A review of the manufacturing documentation did not identify any issues associated with the complaint issue.Labeling was reviewed and sufficiently addresses the customer¿s issue.Based on the investigation, no systemic issue or deficiency of the architect c16000 processing module for serial (b)(6) or the washer, cuvette 16 (rohs) were identified.
|
| |
|