| Model Number |
8667-20 |
| Medical Device Problem Code |
Infusion or Flow Problem (2964)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
09/17/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Information was received from multiple sources (manufacturer representative, healthcare provider) regarding a patient who was receiv ing unknown baclofen via an implantable pump for intractable spasticity.It was reported that the patient was displaying withdrawal symptoms and a dye study was completed and failed.The physician decided to implant an entire new catheter which resolved the issue.
|
| |
|
Additional Manufacturer Narrative
|
|
Section d references the main component of the system.Other medical products in use during the event include: brand name ascenda; product id 8780 (serial: (b)(6)); product type: 0184-catheter; implant date (b)(6) 2020; explant date (b)(6) 2026 medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|